1. Equipment Model
|
|
Coding
|
CT-4008Q-5V50mA-HWX
|
|
2. Indicator Items
|
Index parameters
|
|
Input power supply
|
AC 220V ± 10% / 50Hz
|
|
Input active power
|
20W
|
|
Resolution (AD/DA
|
AD: 24bit; DA: 16bit
|
|
Input impedance
|
≥ 1GΩ
|
|
Energy-saving efficiency
|
/
|
|
Noise
|
< 55dB(Measurement 1 meter away)
|
|
Parallel function
|
Channel parallel connection is not supported
|
|
IP protection level
|
Protection grade IP20
|
|
Channel characteristics
|
The constant current source and the constant voltage source adopt a double closed-loop structure
|
|
Channel control mode
|
Independent control
|
|
Voltage and current detection sampling
|
Four-wire connection
|
|
3. Voltage
|
|
Measurement range
|
10mV~5V
|
|
Minimum discharge voltage
|
-5V
|
|
Precision
|
± 0.01% of FS
|
|
Stability
|
0.02% of FS
|
|
4. Current
|
|
Measurement range
|
Range one: 0.1mA;Range two: 1mA;Range three: 10mA;Range four: 50mA
|
|
Precision
|
± 0.01% of FS
|
|
Minimum output current
|
0.2 μA
|
|
Constant voltage cut-off current
|
Range one: 0.1 μA;Range two: 1 μA Range three: 10μA;Range four: 50μA
|
|
Stability
|
0.02% of FS
|
|
5. Power
|
|
Maximum output power of single channel
|
0.25W
|
|
Precision
|
±0.02% of FS
|
|
Stability
|
0.04% of FS
|
|
6. Time
|
|
Current rise time
|
≤ 1ms (10% ~ 90% of FS)
|
|
Charge and discharge conversion time
|
≤40ms (-90% ~ 90% of FS)
|
|
Working step time range
|
≤(365*24)hour/working steps; Time format support 00 :00 :00.000(h: min: s.ms)
|
|
Sampling frequency
|
10Hz(Minimum time interval: 100ms)
|
|
7. Operating mode
|
|
Charge and discharge mode
|
Constant current charging and discharging, constant voltage charging and discharging, constant current and constant voltage charging and discharging, constant power charging and discharging, pulse charging and discharging, etc
|
|
Cut-off condition
|
Voltage, current, relative time, capacity, energy, etc
|
|
DCIR test
|
Support
|
|
Pulse charge and discharge mode
|
Constant current mode, constant power mode
|
|
Minimum pulse width
|
500ms
|
|
The number of individual pulses
|
A single pulse working step supports 32 different pulses
|
|
Pulse switching
|
One pulse working step can achieve continuous switching of charging or discharging
|
|
Pulse cut-off condition
|
Voltage, relative time
|
|
8. Cycle
|
|
Cycle test range
|
1~65535 times
|
|
Single cycle working steps
|
≤255
|
|
Cycle nesting
|
< 10 layers
|
|
9. Protection
|
|
Software protection
|
Power-off data protection, offline testing function, and safety protection conditions can be set. The set parameters include: upper voltage limit, lower voltage limit, upper current limit, lower current limit, etc
|
|
Hardware protection
|
Reverse connection protection, overvoltage protection, overcurrent protection, over-temperature protection, etc
|